Impact Tester

Scopeof Calibration – Impact Tester

1.TorqueCapacity.
Parameter/ Range / Item
BestMeasurement Capability
(Torqueverification.)
1.0- <10.0 N.m
+/-0.1 %
10.0- <20.0 N.m
+/-0.1 %
20.0- <30.0 N.m
+/-0.1 %
Standard/ Technique / Method /Equipment / Remark
In-house method WI 5.4-13 by direct measument witch Electronicbalance..
Parameter/ Range / Item
BestMeasurement Capability
(PeriodVerification.)
-all hammer
+/-0.7 %
Standard/ Technique / Method /Equipment / Remark
In-house method WI 5.4-14 by direct comparison to standard stop watch.
Parameter / Range / Item
BestMeasurement Capability
(Arm moment verification)
150 -600 mm.
+/-1.30 %
Standard/ Technique / Method /Equipment / Remark
-In-housemethod WI 5.4-14 by direct comparison to standardstop watch.
2.Dimensions
Parameter/ Range / Item
BestMeasurement Capability
(Distance’s support sample.)
20.0- 60.0 mm.
+/-0.01 %.
Standard/ Technique / Method /Equipment / Remark
-In-house method WI 5.4-15 by direct comparison to standard Vernier
Parameter / Range / Item
BestMeasurement Capability
(Heigh of hammer)
200.0- 600.0 mm.
+/-0.25.%
Standard / Technique / Method /Equipment / Remark
-In-house method WI 5.4-15 by direct comparison to standard Vernier.
Parameter / Range / Item
BestMeasurement Capability
(Horizantal level verification)
0.0- 150.0 dreegree.
+/-0.01 dreegree.
Standard / Technique / Method /Equipment / Remark
-In-house method WI 5.4-16 direct measurment by angle gauge.
*expressed as an uncertainty(+/-) which for k=2,providing a lavel of confidence of approximately 95%