Impact Tester
Scopeof Calibration – Impact Tester |
|
![]() |
|
1.TorqueCapacity. | |
Parameter/ Range / Item
|
BestMeasurement Capability
|
(Torqueverification.) | |
1.0- <10.0 N.m
|
+/-0.1 %
|
10.0- <20.0 N.m
|
+/-0.1 %
|
20.0- <30.0 N.m
|
+/-0.1 %
|
Standard/ Technique / Method /Equipment / Remark
|
|
In-house method WI 5.4-13 by direct measument witch Electronicbalance.. | |
Parameter/ Range / Item
|
BestMeasurement Capability
|
(PeriodVerification.)
|
|
-all hammer
|
+/-0.7 %
|
Standard/ Technique / Method /Equipment / Remark
|
|
In-house method WI 5.4-14 by direct comparison to standard stop watch. | |
Parameter / Range / Item |
BestMeasurement Capability
|
(Arm moment verification) | |
150 -600 mm.
|
+/-1.30 %
|
Standard/ Technique / Method /Equipment / Remark
|
|
-In-housemethod WI 5.4-14 by direct comparison to standardstop watch. | |
2.Dimensions | |
Parameter/ Range / Item |
BestMeasurement Capability
|
(Distance’s support sample.) | |
20.0- 60.0 mm.
|
+/-0.01 %.
|
Standard/ Technique / Method /Equipment / Remark
|
|
-In-house method WI 5.4-15 by direct comparison to standard Vernier | |
Parameter / Range / Item |
BestMeasurement Capability
|
(Heigh of hammer) | |
200.0- 600.0 mm.
|
+/-0.25.%
|
Standard / Technique / Method /Equipment / Remark | |
-In-house method WI 5.4-15 by direct comparison to standard Vernier. | |
Parameter / Range / Item |
BestMeasurement Capability
|
(Horizantal level verification) | |
0.0- 150.0 dreegree.
|
+/-0.01 dreegree.
|
Standard / Technique / Method /Equipment / Remark | |
-In-house method WI 5.4-16 direct measurment by angle gauge. | |
*expressed as an uncertainty(+/-) which for k=2,providing a lavel of confidence of approximately 95% |